User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Daytronic 10A72-2:Strain Gage Card DATE: 28-May-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 5 NUMBER OF LINES: 100 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- F 1.002 ASK+ X 1.003 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.004 DISP The following Standard test equipment is required: 1.004 DISP 1. Measurements Group 1550A Strain Indicator Calibrator 1.004 DISP ÿ DTB # 55-2 1.004 DISP 2. Strain Indicator Calibration Cable 1.005 DISP Connect the Daytronic Mainframe to the system Com1. 1.006 DISP With the Daytronic Model 10K6 mainframe turned off, 1.006 DISP remove any conditioner card that may be located in 1.006 DISP slot #1. 1.007 DISP Install the UUT into slot #1. 1.008 DISP Connect a Daytronic CONDITIONER CONNECTER model 60322, 1.008 DISP wired for two channels, to the rear of the UUT. 1.009 DISP Power up the Model 10K6 mainframe. 1.010 HEAD {} 1.011 JMP 1.018 1.012 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.013 HEAD {º EXCITATION VOLTAGE TEST º} 1.014 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.015 HEAD {}ÍÍ EXCITATION VOLTAGE TEST ÍÍ 1.016 DISP ÿ Connect the UUT and the 1281 as follows: 1.016 DISP ÿ[27][91]1m CHANNEL 1 TO 1281[27][91]m 1.016 DISP ÿ OUTPUT (red) ÄÄÄÄÄÄÄÄÄÄÄÄ> INPUT-HIGH 1.016 DISP ÿ OUTPUT (black) ÄÄÄÄÄÄÄÄÄÄÄÄ> INPUT-LO 1.017 CALL Daytronic Excitation Sub 1.018 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.019 HEAD {º Excitation Voltage Test º} 1.020 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.021 HEAD {ÍÍ Channel 1 ÍÍ} 1.022 HEAD ÍÍ Excitation Voltage Test ÍÍ 1.023 JMP 2.001 1.024 EVAL 2.001 DISP ÿÿÿÿÿÿÿÿConnect the UUT as follows: 2.001 DISP + EXCITATION (pin 1) ÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]1281 INPUT HI 2.001 DISP - EXCITATION (pin 4) ÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]1281 INPUT LO 2.002 CALL Daytronic Excitation Sub 2.003 HEAD {} 2.004 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 2.005 HEAD {º Strain Indication Tests º} 2.006 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 2.007 HEAD {ÍÍ Channel 1 ÍÍ} 2.008 HEAD ÍÍ Channel 1 Strain Indication Tests ÍÍ 2.009 JMP 3.001 2.010 EVAL 3.001 DISP Set the 1550A 350ê bridge S+ shorting link to the closed 3.001 DISP position. 3.002 DISP ÿ Connect the UUT channel 1 as follows: 3.002 DISP +SIGNAL (pin 2) ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]1550A +S 3.002 DISP -SIGNAL (pin 3) ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]1550A -S 3.002 DISP + EXCITATION (pin 1) ÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]1550A +P 3.002 DISP - EXCITATION (pin 4) ÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]1550A -P 3.003 PORT RST[13] 3.004 PORT BIT999=1[13] # 3.005 PORT LCT 1=1011[13] 3.005 PORT LCT 1=11[13] 3.006 PORT RST 1[13] 3.007 PORT TYP 1=72[13] 3.008 MATH m[1]=1 3.009 CALL Daytronic Strain Indication Sub 3.010 HEAD {} 3.011 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 3.012 HEAD {º Excitation Voltage Test º} 3.013 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 3.014 HEAD {ÍÍ Channel 2 ÍÍ} 3.015 HEAD ÍÍ Excitation Voltage Test ÍÍ 3.016 JMP 4.001 3.017 EVAL 4.001 DISP Move the connector from Channel 1 to Channel 2. 4.002 CALL Daytronic Excitation Sub 4.003 HEAD {} 4.004 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 4.005 HEAD {º Strain Indication Tests º} 4.006 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 4.007 HEAD {ÍÍ Channel 2 ÍÍ} 4.008 HEAD ÍÍ Channel 2 Strain Indication Tests ÍÍ 4.009 JMP 5.002 4.010 EVAL 5.001 PORT BIT999=1[13] # 5.002 PORT LCT 2=1012[13] 5.001 PORT LCT 2=12[13] 5.001 PORT TYP 2=72[13] 5.002 MATH m[1]=2 5.003 CALL Daytronic Strain Indication Sub 5.004 HEAD {} 5.005 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 5.005 DISP {º THIS COMPLETES THE VERIFICATION º} 5.005 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 5.006 END